| INPUT |
| Voltage |
115/230 VAC ± 10%, User selection |
| Frequency |
50/60 Hz ± 5% |
| Fuse |
6.3 Amp 250V Slo-Blo |
| DIELECTRIC WITHSTAND TEST MODE |
| Output Rating |
5 KV @ 20mA AC |
| 6 KV @ 5 mA DC |
| Output Adjustment |
Range: |
0-5 KV AC, 0-6 KV DC |
| Resolution: |
10 volt/step |
| Accuracy: |
± (2% of setting + 5 volts) |
| Voltage Display |
Range: |
0 - 5 KVAC / 0-6KV DC |
| Resolution: |
10 volt/step |
| Accuracy: |
± (2% of reading + 10 volts) |
| HI / LO -Limit AC |
Range: |
0.00 - 20.00 mA |
| Resolution: |
0.01 mA/step |
| Accuracy: |
± (2% of setting + 0.02 mA) |
| AC Current Display |
Range: |
0.00 - 20.00 mA |
| Resolution: |
0.01 mA/step |
| Accuracy: |
± (2% of reading + 0.02 mA) |
| HI / LO -Limit DC |
Range: |
0.00 - 5.00 mA |
| Resolution: |
0.01 mA/step |
| Accuracy: |
± (2% of setting + 0.02 mA) |
| DC Current Display |
Range: |
0.00 - 5.00 mA |
| Resolution: |
0.01 mA/step |
| Accuracy: |
± (2% of reading + 0.02 mA) |
| Failure Detector |
Audible and Visual |
| Output Ripple |
< 5% (6KVDC/5 mA at Resistive Load) |
| Discharge Time |
< 200 ms |
| Maximum Capacitive Load in DC Mode |
1µF < 1KV , 0.75µF < 2KV, 0.5µF < 3KV, 0.08µF < 4KV, 0.04µF < 5KV, 0.01µF < 6KV |
| AC Wave Form |
Sine wave Distortion : < 2% THD ; Crest Factor= 1.3 – 1.5 |
| Output Frequency |
Range: |
60 or 50 Hz, User Selection |
| Output Regulation |
± (1 % of output + 5 volts) from no load to full load |
| Dwell Timer |
Range: |
0, 0.2 - 999.9 sec, (0 = Constant) |
| Resolution: |
0.1 sec increments |
| Accuracy: |
± (0.1% + 0.05 sec) |
| Ramp Timer |
Range: |
0.1 - 999.9 sec |
| Resolution: |
0.1 sec increments |
| Accuracy |
± (0.1% + 0.05 sec) |
| INSULATION RESISTANCE TEST MODE |
| Output Voltage |
Range: |
100 - 1000 Volts DC |
| Resolution: |
1 volt/step |
| Accuracy: |
± (2% of reading + 5 volts) |
| Voltage Display |
Range: |
0 - 1000 V |
| Resolution: |
1 volt/step |
| Accuracy: |
± (2% of reading + 2 counts) |
| Resistance Display |
Range: |
1 - 1000 MΩ (4 Digit, Auto Ranging) |
| Resolution: |
0.01/0.1 MΩ |
| Accuracy: |
± (3% of reading + 2 counts) at test voltage > 500V |
| ± (7% of reading +2 counts) at test voltage < 500V |
| HI-Limit |
Range: |
0, 1 - 1000 M Ω (0 = Off) |
| LO-Limit |
Range: |
1 - 1000 M Ω |
| Delay Timer |
Range: |
0, 0.5 - 999.9 sec (0 = Constant) |
| Resolution: |
0.1 sec/step |
| Accuracy: |
± (0.1% + 0.05 sec) |
| GROUND BOND TEST MODE |
| Output Voltage |
Range: |
AC 6 volts Max. |
| Output Frequency |
Range: |
60 or 50 Hz, User Selection |
| Accuracy: |
±0.1% |
| Output Current |
Range: |
3.0 - 30.0 Amps AC |
| Resolution: |
0.1 Amp/step |
| Accuracy : |
± (2 % of setting + 0.1A) |
| Current Display |
Range: |
0.0 - 30.0 Amps |
| Resolution: |
0.1 Amp/step |
| Accuracy: |
± (3 % of reading + 0.1A) |
| Resistance HI-Limit |
Range: |
0 - 510 mΩ |
| Resistance Display |
Range: |
0 – 510 mΩ for 0.0 – 10.0 Amps |
| 0 – 200 mΩ for 10.1 – 25.0 Amps |
| 0 – 150 mΩ for 25.1 – 30.0 Amps |
| Resolution: |
1 mΩ/step |
| Accuracy: |
± (2 % of Setting + 2 mΩ ) |
| Dwell Timer |
Range: |
0, 0.5 - 999.9 sec (0 = Constant) |
| Resolution: |
0.1 sec/step |
| Accuracy: |
± (0.1% + 0.05 sec) |
| Lead Resistance Offset |
Max. Offset Capability: |
100 mΩ |
| Resolution: |
1 mΩ / step |
| Accuracy: |
± (2 % of Setting + 2 mΩ ) |
| GENERAL SPECIFICATIONS |
| PLC Remote Control |
Input - Test, Reset, Interlock & Recall Memory 1 - 6 |
| Output - Pass, Fail, Test-in-Process |
| Memory |
Allows storage of up to 6 different test programs and 6 step/each memory and single step mode. |
| Security |
Lockout capability to avoid unauthorized access to test set-up programs. |
| Calibration |
Software and adjustments are made through front panel. |
| Mechanical |
Bench or rack mount with tilt up front feet. |
| Optional Arc Detection Hardware |
ACW and DCW; Software adjustable Arc Level 0-9 |
| Dimension |
280(W) x 89(H) x 430(D) mm |
| Weight |
33 lbs. (15 kg) |