| INPUT |
| Voltage |
115 / 230Vac ± 15% |
| Frequency |
50 / 60Hz ± 5% |
| Fuses |
Fuse 6.3A / 250V Slow-Blow |
| |
Range
|
Resolution
|
Accuracy
|
| AC WITHSTAND VOLTAGE |
|
| Output Voltage, KVAC |
0-4.00 |
0.01 |
± (2% of setting + 5V) |
| Output Frequency |
50Hz/60Hz ±0.1%, User Selection |
| Output Waveform |
Sine Wave , Crest Factor = 1.3 - 1.5 |
| Output Regulation |
± (1% of output + 5V), From no load to full load |
| SETTINGS |
| HI and LO-Limit current, mA |
0.00-40.00 |
0.01 |
± (2% of setting + 2 counts) |
| Ramp Timer, second |
0.1 - 999.9 |
0.1 |
± (0.1% of setting + 0.05 sec) |
| Dwell Timer, second |
0, 0.2 - 999.9
(0=continuous) |
| Current Offset, mA |
0.00-2.00 |
0.01 |
± (2% of setting + 2 counts) |
| DUT Output |
ON/OFF BY User Selection |
| Arc Detection(Option) |
1-9 ranges (9 is the most sensitivity) |
| DC WITHSTAND VOLTAGE |
| Output Voltage, KVDC |
0 -4.00 |
0.01 |
± (2% of setting + 5V) |
| DC Output Ripple |
<5 % at 4KV/5mA |
| SETTINGS |
| HI and LO-Limit current, mA |
0.00 - 5.00 |
0.01 |
± (2% of setting + 2 counts) |
| Ramp Timer, second |
0.1- 999.9 |
0.1 |
± (0.1% of setting + 0.05 sec) |
| Dwell Timer, second |
0, 0.2 - 999.9
(0=continuous) |
| Current Offset, mA |
0.00 - 0.20 |
0.01 |
± (2% of setting + 2 counts) |
| DUT Output |
ON/OFF BY User Selection |
| Arc Detection(Option) |
1-9 ranges (9 is the most sensitivity) |
| Discharge Time |
No capacitive Load is < 30 msec, capacitive Load is < 100 msec. |
| Maximum Capacitive Load DC Mode |
1µF < 1KV,0.5µF < 3KV,0.75µF < 2KV, 0.08µF < 4KV |
| INSULATION RESISTANCE |
| Output Voltage, DCV |
100 - 1000 |
1 |
± (2% of reading + 5 volts) |
| SETTINGS |
| HI-Limit, resistance, MΩ |
0,1 - 1000 (0 = OFF) |
1 |
Same as Resistance MEASUREMENT Accuracy |
| Lo-Limit, resistance, MΩ |
1 - 1000 |
| Delay Timer, second |
0, 0.5 - 999.9
(0=continuous) |
0.1 |
± (0.1% of setting + 0.05 sec) |
| GROUND BOND |
| Output AC Current, A |
3.0 – 30.0 |
0.1 |
± (2 % of setting + 0.02A) |
| Output Voltage, Vac |
AC 6 volts Max. |
| Output Frequency, Hz |
50Hz/60Hz ±0.1%, User Selection |
| Milliohm Offset |
0-100 |
1 |
± (2 % of setting + 2 counts) |
| SETTINGS |
| HI and LO-Limit Resistance, mΩ |
0 – 510 (3.0-10.0A) |
1 |
± (2 % of setting + 2 counts ) |
| 0 – 200 (10.1-25.0A) |
| 0 – 150 (25.1-30.0A) |
| Dwell Timer, second |
0, 0.5 - 999.9
(0 = continuous) |
0.1 |
± (0.1% of setting + 0.05 sec) |
| MEASUREMENT |
| |
Range |
Resolution |
Accuracy |
| Voltage, KV(AC/DC) |
0.00 - 4.00 |
0.01 |
± (1.5 % of reading +1 count) |
| Voltage, Vdc (IR only) |
0 - 1000 |
1 |
± (1.5% of reading + 2 counts) |
| AC Current, mA |
0.00 - 40.00 |
0.01 |
± (2% of reading + 2 counts) |
| DC Current, mA |
0.00 – 5.00 |
0.01 |
| AC Current, A (G.B) |
0.0-30.0 |
0.1 |
± (3 % of reading + 3 counts) |
| Resistance, MΩ (IR) |
1-999.9 |
0.1/0.01 |
100-499V ± (7% of reading + 2 counts)
500-1000V ± (3% of reading + 2 counts) |
| Resistance, mΩ (GB) |
0-510 |
1 |
± (2 % of reading + 2 counts) |
| RUN TEST |
| DUT POWER |
| AC Voltage |
0 - 277.0V, Single phase unblance, 0 - 30.0A maximum |
| Current |
30A maximun continuous |
| Power Rating |
8400W maximum |
| DUT Protection |
Short Circuit current 50A < 3 sec, Inrush Current 180A Response time 10u sec. |
| SETTINGS |
| HI and LO Limit AC Voltage, V |
0.0-277.0 |
0.1 |
± (1.5% of setting + 0.2 V), 60.0-277V |
| HI and LO Limit AC Current, A |
0.0-30.0 |
0.1 |
± (2% of setting + 2 counts) |
| HI and LO Limit AC Power, W |
0 - 8400 |
1 |
± (5% of setting + 6 counts) |
| HI and LO Limit Power Factor |
0.000-1.000 |
0.001 |
W/VA, calculated and displayed to three significant digits |
| HI and LO Limit Leakage Current |
0.0-010.00, Hi-Limit
0=OFF |
0.01 |
± (2% of setting + 2 counts) |
| Delay Time, second |
0.2-999.9 |
0.1 |
± (0.1% of setting + 0.05 sec) |
| Dwell Time, second |
0, 0.1-999.9 (0=continuous) |
| MEASUREMENT |
| |
Range |
Resolution |
Accuracy |
| Voltage, Vac |
0.0-277.0 |
0.1 |
± (1.5% of reading + 2counts), 60.0~277V |
| Current, Aac |
0.0-30.0 |
0.1 |
± (2% of reading + 2counts) |
| Power, Watts |
0-8400 |
1 |
± (5% of reading + 6 counts) |
| Power, Factor |
0.000-1.000 |
0.001 |
W/VA, calculated and displayed to three significant digits |
| Leakage Current, mA |
0.00-10.00 |
0.01 |
± (2% of reading + 2 counts) |
| MD Circuit |
Leakage Current measuring resistance = 2K Ω ±1% |
| Timer, second |
0.0-999.9 |
0.1 |
± (0.1% of reading + 0.05 sec) |
| TOUCH CURRENT TEST |
| DUT |
| DUT Input Power Rating |
0 - 277.0V, Single phase unblance, 30A maximum |
| Current |
30A maximun continuous |
| DUT Protection |
Short Circuit current 50A < 3 sec, Inrush Current 180A Response time 10u sec. |
| SETTINGS |
| HI and LO Limit, mA |
Range 0.00-10.00mA, Resolution 0.01mA,Hi-Limit 0=OFF
|
| Delay Time, second |
Range 0, 1.0-999.9 (0=continous), Resolution 0.1second.
|
| Compliance Standards |
A UL544 NP
|
|
B IEC60990 Fig4 U2
|
|
X External MD(1K Ohm)
|
|
F Frequency check, External MD(1K Ohm)
|
| Line Condition |
NEUTRAL, REVERSE, GROUND
|
| Probe settings |
G-L, G-N
|
| CURRENT MEASUREMENT |
| Frequency Range |
DC-100KHz |
| Touch current Range (RMS) |
Resolution |
Accuracy (DC,15-100KHz) |
| 0.00mA - 10.00mA |
0.01 |
±(2% of reading + 3counts) |
| Touch current offset |
Resolution |
Accuracy (DC,15-100KHz) |
| 0.00mA - 2.00mA |
0.01 |
±(2% of reading + 3counts) |
| LINE VOLTAGE MEASUREMENT |
| Range |
0.0-277.0 VAC |
| Resolution |
0.1 V |
| Accuracy |
± (1.5% of reading +0.2V), 60.0-277.0V |
| External MD |
Basic measuring element 1K Ω |
| MD A - D components |
Accuracy: Resistance ± 1% Capacitance ± 5% |
| MD Voltage Limit |
Maximum 30V peak or 30VDC |
| External Terminal on TC, allowing for High, Low, Open settings (Option) |
Add a extemal chanel to reach P-G, S-G and P-S for hipot test. |
| |
Extemal chanel |
HV relay |
GND relay |
| P-G |
Open |
ON |
ON |
| S-G |
High |
OFF |
ON |
| P-S |
Low |
ON |
OFF |
| GENERAL |
| PLC Remote Control |
Input : Test, Reset, Interlock, Recall Memory 1-6
1.Minimun 5ms Plus time for Remote Input key.
2.It need 100ms for test key - output (ACW,DCW,IR,GB). RUN and LLT need 200ms.
3.It need 100ms for reset key - output shutdown.
4.It need 100ms for interlock - output shutdown.
5.Minimun stable data time on Memory Select input need 300ms.
6.Maximun 2ms between Memory bit changes.
|
|
Output:- Pass, Fail, Test-in-Process
1.Output relay contact ratings is 1A/125V
2.The output is separation output
|
| Memory |
20 memories, 10 steps, step and step can link.
RUN and LLT step by step need 1.S, other need 100ms. Single step mode
|
| Remote Output |
provide the +24V, GND and I/O1~7 to control 67xx and 68xx AC Power Source or Transformer Box 1931
|
| Display |
20 x 2 LCD
|
| LED display |
BUS REM(WITH RS232), LOCK, NEUTRAL, REVERSE, GROUND
|
| Interface |
Optional RS232
|
| Security |
Key Lock capability to avoid unauthorized access to all test parameters. Memory Lock capability to avoid unauthorized access to memory locations.
|
| Fail stop |
This allows the 6330 to be configured to continue testing after detecting a failure.All pass/fail results are displayed at the completion of the test cycle.
|
| Alarm Volume Setting |
Range : 0~9 ; 0=OFF, 1 is softest volume, 9 is loudest volume.
|
| Calibration |
Software and adjustments are made through the front panel.
|
| Environment |
0-40 ℃ , 20-80%RH
|
| Dimensions/Net Weight |
430mm(W)×133mm(H)×500 mm(D)/22Kg
|
| STANDARD ACCESSORIES |
| Power Cord(10A) |
x1 |
| Fuses |
×2 (Including a spare contained in the fuse holder) |
| Interlock Disable Key(1505) |
x1 |
| Hipot Test Lead, 1.5m (1101) |
x1 |
| Ground Bond Test Lead 40A, 3m(1153) |
x1 |
| Ground Bond Return Lead 40A, 3m(1152) |
x1 |
| LLT Receptacle Adaptor Box 30A , 1.6m(1932) |
x1 |
| DUT Power Input Lead 40A, 3m(1151) |
x1 |
| DUT Power Output Lead 40A, 3m(1150) |
x1 |